Product Regulations
2
The unit passed with criterion A except for the probe ID pads on the inputs. A discharge to these pads caused the software
to lock up requiring operator intervention. Hence, the ID pads are rated at C.
EMC
Performance Criteria
1
IEC 61326-1:1997+A1:1998/EN 61326-1:1997+A1:1998
CISPR 11:1990 / EN 55011:1991
IEC 61000-4-2:1995+A1:1998/EN 61000-4-2:1995 A
2
IEC 61000-4-3:1995/EN 61000-4-3:1995 A
IEC 61000-4-4:1995/EN 61000-4-4:1995 A
IEC 61000-4-5:1995/EN 61000-4-4:1995 A
IEC 61000-4-6:1996/EN 61000-4-6:1996 B
IEC 61000-4-11:1994/EN 61000-4-11:1994 A
Canada: ICES-001:1998
Australia/New Zealand: AS/NZS 2064.1
Safety IEC 61010-1:1990+A1:1992+A2:1995/EN 61010-1:1994+A2:1995
Canada: CSA-C22.2 No. 1010.1:1992
USA: UL 3111-1:1994 {optional}
Additional Information:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC
and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE-marking accordingly
(European Union).
1
Performance Criteria:
A Pass - Normal operation, no effect.
B Pass - Temporary degradation, self recoverable.
C Pass - Temporary degradation, operator intervention required.
D Fail - Not recoverable, component damage.
Sound Pressure
Level
Less than 60 dBA
Regulatory Information for Canada
ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est confomre à la norme NMB-001 du Canada.
Regulatory Information for Australia/New Zealand
This ISM device complies with Australian/New Zealand AS/NZS 2064.1