General Troubleshooting 2
Troubleshooting and Maintenance Manual 31
Peaks have split tops
Bad injection technique
Injection is too large
Baseline is rising
Column bleed
Other contamination
Baseline is high
Column bleed
Other contamination
Baseline is falling
A falling baseline indicates contamination is being swept away. Wait until the
baseline reaches an acceptable level. Common causes include:
Residual air and water from a recent venting
Column bleed
Septum bleed
Splitless injection time too long (inlet is not properly swept, resulting in
excess solvent on the column and slow solvent decay)