60
Using the Agilent 81101A
The Output Screen
NOTE
The leading and trailing edges are independently programmable within
certain ranges only, see Transition Times in the Reference Guide.
NOTE
You cannot have the width format set to D
TY
C
YC
and the leading/trailing-
edge format set to percentage of width (L
EAD
E%/T
RAI
E%) at the same
time.
Pulse Trailing Edge Parameter
Set the trailing edge transition-time of the pulse, measured from 10% to
90% of the pulse amplitude. There are three formats available:
=L
EAD
E (select =L
EAD
E)
The trailing edge transition time is coupled directly to the leading
edge to maintain a symmetrical pulse.
T
RAI
E (select A
BSOLUTE
)
The absolute transition time measured from 90% to 10% of the pulse
amplitude. In this format the trailing edge is independent of the
leading edge and the pulse width.
T
RAI
E% (select %
OF
W
ID
)
The trailing edge transition time expressed as a percentage of the
pulse width. In this format, if you adjust the pulse width, the
transition time is adjusted to maintain the edge time as a percentage
of the width.
NOTE
The leading and trailing edges are independently programmable within
certain ranges only, see Transition Times in the Reference Guide.
NOTE
You cannot have the width format set to D
TY
C
YC
and the leading/trailing
edge format set to percentage of width (L
EAD
E%/T
RAI
E%) at the same
time.