The FREF command specifies the frequency reference source. Select
either the internal frequency reference (INT) or supply your own
external reference (EXT). An external reference must be 10 MHz (±100
Hz) at a minimum amplitude of 0 dBm. Connect the external reference
to J9 (10 MHz REF IN/OUT) on the rear panel. When the external
mode is selected, an "X" appears on the left edge of the display.
Preset State
Internal
Query Response
Figure 7-119 FREF Query Response
Example
10 INPUT "WHAT IS THE FREQUENCY REFERENCE SOURCE (INT OR
EXT)",Src$
20 OUTPUT 718;"FREF ";Src$;";"
30 PRINT "SOURCE SELECTED IS",Src$
40 END