Chapter 7 493
Language Reference
IDCF Signal Identification to Center Frequency
IDCF Signal Identification to Center
Frequency
Syntax
Figure 7-140 IDCF Syntax
Description
The IDCF command sets the center frequency to the frequency obtained
from the command SIGID. SIGID must be in AUTO mode and have
found a valid result for this command to execute properly. Use SIGID
on signals when in external mixing mode. IDCF only applies to
spectrum analyzers with firmware revisions 920528 or with Option
008. The command does not apply to an 8560E/EC Option 002 or when
the spectrum analyzer is configured to use preselected external mixers.
Example
10 OUTPUT 718;"SIGID AUTO;"
20 OUTPUT 718;"IDCF;"
30 OUTPUT 718;"CF?;"
40 ENTER 718;Cf
50 PRINT Cf
60 END