EasyManuals Logo

Agilent Technologies 8562EC User Manual

Agilent Technologies 8562EC
716 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #606 background imageLoading...
Page #606 background image
606 Chapter7
Language Reference
SRCPSTP Source Power Step
SRCPSTP Source Power Step
Syntax
Figure 7-253 SRCPSTP Syntax
Description
The SRCPSTP command sets the step size of the source power level,
source power offset, and power-sweep range functions. This function is
available only with an 8560E/EC Option 002.
Parameters
number real from 0.1 dB to 12.75 dB; 0.05 dB resolution via
GPIB.
UP or DN 0.1 dB steps.
Preset State
1.0 dB
Query Response
Figure 7-254 SRCPSTP Query Response

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8562EC and is the answer not in the manual?

Agilent Technologies 8562EC Specifications

General IconGeneral
BrandAgilent Technologies
Model8562EC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals