EasyManuals Logo

Agilent Technologies 8720ES Reference Guide

Agilent Technologies 8720ES
263 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #175 background imageLoading...
Page #175 background image
6-4
Error Messages
Error Messages in Alphabetical Order
ARGUMENT OUT OF RANGE
Error Number
206
The argument for a programming command is out of the specified range. Refer
to the programmer’s guide for a list of programming commands and argument
ranges.
ASCII: MISSING 'BEGIN' STATEMENT
Error Number
193
The CITIfile you just downloaded over the GPIB or via disk was not properly
organized. The analyzer is unable to read the “BEGIN” statement.
ASCII: MISSING 'CITIFILE' STATEMENT
Error Number
194
The CITIfile you just downloaded over the GPIB or via disk was not properly
organized. The analyzer is unable to read the “CITIFILE” statement.
ASCII: MISSING 'DATA' STATEMENT
Error Number
195
The CITIfile you just downloaded over the GPIB or via disk was not properly
organized. The analyzer is unable to read the “DATA” statement.
ASCII: MISSING 'VAR' STATEMENT
Error Number
196
The CITIfile you just downloaded over the GPIB or via disk was not properly
organized. The analyzer is unable to read the “VAR” statement.
AVERAGING INVALID ON NON-RATIO MEASURE
Error Number
13
You cannot use sweep-to-sweep averaging in single-input measurements.
Sweep-to-sweep averaging is valid only for ratioed measurements (A/R, B/R,
A/B, and S-parameters). You can use other noise reduction techniques, such as
narrower IF bandwidth, for single input measurements.
BAD FREQ FOR HARMONIC OR FREQ OFFSET
Error Number
181
You turned on time domain or recalled a calibration that resulted in start and
stop frequencies that are beyond the allowable limits.
BANDWIDTH LIMIT INVALID: MIN BW > MAX BW
Information
Message
The bandwidth test has a minimum bandwidth greater than the maximum
bandwidth. Change the minimum bandwidth to a frequency less than the
frequency of the maximum bandwidth or change the maximum bandwidth to a
frequency greater than the frequency of the minimum bandwidth

Table of Contents

Other manuals for Agilent Technologies 8720ES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720ES and is the answer not in the manual?

Agilent Technologies 8720ES Specifications

General IconGeneral
BrandAgilent Technologies
Model8720ES
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals