Page #176 background imageLoading...
Page #176 background image
6-5
Error Messages
Error Messages in Alphabetical Order
BATTERY FAILED. STATE MEMORY CLEARED
Error Number
183
The battery protection of the non-volatile CMOS memory has failed. The CMOS
memory has been cleared. Refer to the service guide for battery replacement
instructions. See Chapter 8 , “Preset State and Memory Allocation” for more
information about the CMOS memory.
BATTERY LOW! STORE SAVE REGS TO DISK
Error Number
184
The battery protection of the non-volatile CMOS memory is in danger of failing.
If this occurs, all of the instrument state registers stored in CMOS memory will
be lost. Save these states to a disk and refer to the service guide for battery
replacement instructions. See Chapter 8 , “Preset State and Memory Allocation”
for more information about the CMOS memory.
BLOCK INPUT ERROR
Error Number
34
The analyzer did not receive a complete data transmission. This is usually
caused by an interruption of the bus transaction. Clear by pressing the
key or aborting the I/O process at the controller.
BLOCK INPUT LENGTH ERROR
Error Number
35
The length of the header received by the analyzer did not agree with the size of
the internal array block. Refer to the programmer’s guide for instructions on
using analyzer input commands.
CALIBRATION ABORTED
Error Number
74
You have changed the active channel during a calibration, so the calibration in
progress was terminated. Make sure the appropriate channel is active and
restart the calibration.
CALIBRATION REQUIRED
Error Number
63
A calibration set could not be found that matched the current stimulus state or
measurement parameter. You will have to perform a new calibration.
CANNOT DETERMINE ORIENTATION
Information
Message
There is a problem with the ECal module in the RF path. Check RF connections,
power connections, and interface connections to module.
Local

Table of Contents

Other manuals for Agilent Technologies 8720ES

Questions and Answers: