Verifying Specifications
This chapter tells you how to use the Agilent 89410A Auto Performance Test
disk. The performance test disk contains a program that semiautomates the
operation verification tests and performance tests.
After you review this chapter, follow the directions in ‘’To load the program’’
then continue with one of the following:
l
‘’To run the program in semiautomated mode’’
l
‘’To run the program without a printer’’
l
‘’To run the program in manual mode’’
Caution Before applying line power to the analyzer or testing its electrical performance,
see chapter 1, ‘’Preparing the Analyzer for Use.’’
Overview
The Auto Performance Test disk contains a program (ITM_89410A) and two
procedure files (OP_VERIFY and PERFORMAN). ITM_89410A is the test
manager program. OP_VERIFY is the operation verification procedure file and
PERFORMAN is the performance test procedure file. The procedure files
contain an ordered list of tests, and each test contains one or more
measurements. Since ITM_89410A reads the procedure files, the disk must
remain in the disk drive during testing.
The channel 2 BNC connector is installed in both 1 and 2 channel analyzers. To
determine if the optional second channel (option AY7) is installed, press [
System
Utility
][
option setup
].
If you do not have a keyboard connected to the analyzer, use the numeric key
pad and the alpha keys when the program prompts you to type in information.
See the analyzer’s help text for a description of the alpha keys.
If a test fails, contact your local Agilent Technologies sales and service office or
have a qualified service technician see chapter 1, ‘’Troubleshooting the
Analyzer,’’ in the Agilent 89410A Service Guide.
Verifying Specifications Agilent 89410A
2-2