To set up the input resistance test
Performance Test only
This test verifies that the Agilent 89410A meets its input specification for
impedance. In this test, input resistance is measured directly using a digital
multimeter. The 10 M range is used on the digital multimeter to prevent
the current from turning on the overload protection FET in the input
circuitry.
1
2
Option AY7 only
Verifying Specifications Agilent 89410A
To set up the input resistance test
2-32