To install the analyzer
The analyzer is shipped with plastic feet in place, ready for use as a
portable bench analyzer. The plastic feet are shaped to make full-width
modular instruments self-align when they are stacked.
l Install the analyzer to allow free circulation of cooling air.
Cooling air enters the analyzer through the rear panel and exhausts
through both sides.
Warning To prevent potential fire or shock hazard, do not expose the analyzer
to rain or other excessive moisture.
l Protect the analyzer from moisture and temperatures or temperature changes
that cause condensation within the analyzer.
The operating environment specifications for the analyzer are listed in the
Agilent 89410A Technical Data publication.
Caution Use of the equipment in an environment containing dirt, dust, or corrosive
substances will drastically reduce the life of the disk drive and the flexible disks.
The flexible disks should be stored in a dry, static-free environment.
l To install the analyzer in an equipment cabinet, follow the instructions shipped
with the rack mount kits.
Agilent 89410A Preparing the Analyzer for Use
To install the analyzer
1-7

Table of Contents

Other manuals for Agilent Technologies 89410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 89410A and is the answer not in the manual?

Agilent Technologies 89410A Specifications

General IconGeneral
BrandAgilent Technologies
Model89410A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals