3 Press the following keys and type in the printer address when the program
prompts you:
[
TEST CONFIG
]
[
PRINTER ADDRESS
]
[
PROCEDURE
]
[
OP_VERIFY
]or[
PERFORMAN
]
[
STOP AFTER
]
[
LIMIT FAILURE
]or[
NONE
]
[
RETURN
]
4 Press the following keys to start the test:
[
START TESTING
]
[
START BEGINNING
]
When you select [
START BEGINNING
], the data is written to a file on the disk
and printed only after all tests are done. When you select [
START MIDDLE
]
or [
ONE TEST
], the data is printed immediately after each measurement.
5 Follow the directions on the display.
The directions on the display briefly tell you how to connect test
equipment. For detailed illustrations of equipment setup, see the setup
illustrations starting on page 2-14.
If you want to pause the program and return the Agilent 89410A DC-10
MHz Vector Signal Analyzer to front panel control, press [
BASIC
]. To
continue the program, press [
Display
][
BASIC display format
][
lower
][
BASIC
]
[
continue
]. If you changed any instrument setup states, press [
RESTART TEST
]
to ensure accurate measurement results.
Agilent 89410A Verifying Specifications
To run the program in semiautomated mode
2-11

Table of Contents

Other manuals for Agilent Technologies 89410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 89410A and is the answer not in the manual?

Agilent Technologies 89410A Specifications

General IconGeneral
BrandAgilent Technologies
Model89410A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals