EasyManuals Logo
Home>Agilent Technologies>Power Supply>E3633A

Agilent Technologies E3633A Operating Manual

Agilent Technologies E3633A
175 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
Chapter 4 Remote Interface Reference
Status Reporting Commands
100
*PSC?
This command queries the power-on status clear setting. The returned
parameter is ‘‘0
(*PSC 0
) or ‘‘1’’ (
*PSC 1
).
*SRE <enable value>
This com mand enables bits in the Status Byte enable register.
*SRE?
This com mand queries the Status Byte Enable register. The power supply
returns a decimal value which corresponds to the binary-weighted sum of all
bits set in the enable register.
*STB?
This command queries the Status Byte summary register. The
*STB?
command
is similar to a serial poll but it is processed like any other instrument command.
The
*STB?
command returns the sa me result as a serial poll but the ‘‘Request
Service’’ bit (bit 6) is not cleared if a serial poll has occurred.
*WAI
This command instructs the power supply to wait for all pending operations
to complete before executing any additional commands over the interface.
Used only in the triggered mode.

Table of Contents

Other manuals for Agilent Technologies E3633A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3633A and is the answer not in the manual?

Agilent Technologies E3633A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3633A
CategoryPower Supply
LanguageEnglish

Related product manuals