Chapter 4 Remote Interface Reference
An Introduction to the SCPI Language
103
4
Command Separators
A
colon
( : ) is used to separate a command keyword from a lower-level
keyword as shown below :
‘‘
SOURce:CURRent:TRIGgered
’’
A semicolon ( ; ) is used to separate two commands within the same subsystem,
and can also minimize typing. For example, sending the following command
string:
‘‘
SOUR:VOLT MIN;CURR MAX
’’
... is the same as sending the following two commands:
‘‘
SOUR:VOLT MIN
’’
‘‘
SOUR:CURR MAX
’’
Use a colon and a semicolon to link commands from different subsystems. For
example, in the following command string, an error is generated if you do not
use the colon and semicolon:
‘‘
DISP:TEXT:CLE;:SOUR:CURR MIN
’’
Using the MIN and MAX Parameters
You can substi tute
MINimum
or
MAXimum
in place of a parameter for many
commands . For exampl e, consi der the followi ng comm and:
CURRent {<current>|MIN|MAX}
Instead of selecting a specific current, you can substitute
MINimum
to set the
current to its minimum value or
MAXimum
to set the current to its maximu m
value.

Table of Contents

Other manuals for Agilent Technologies E3633A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3633A and is the answer not in the manual?

Agilent Technologies E3633A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3633A
CategoryPower Supply
LanguageEnglish

Related product manuals