Chapter 4 Remote Interface Reference
Calibration Commands
87
4
CALibration:DAC:ERRor
This command c orrects the differential nonlinearity error of the internal DAC
without an external meter. You must send this command before calibrating the
voltage. It takes about 30 seconds to execute the comm and.
CALibration:SECure:CODE <new code>
This command enters a new security code. To change the security code, first
unsecure the power supply using the old security code. Then, enter the new
code. The calibration code may contain up to 12 characters over the remote
interface but the first character must always be a letter.
CALibration:SECure:STATe {OFF | ON},<code>
This command unsecures or secures the power supply for calibration. The
calibration code may contain up to 12 characters over the remote interface.
CALibration:SECure:STATe?
This command queries the secured state for calibration of the power supply.
The returned parameter i s ‘‘0’’ (OFF) or ‘1’’ (ON).
CALibration:STRing <quoted string>
This command records calibration information about your power supply. For
example, you can store such information as the last calibration dat e, the ne xt
calibration due date, or the power supplys serial number. The calibration
message may contain up to 40 characters. The power supply shou ld be
unsecured before sending a calibration message.
CALibration:STRing?
This command queries the calibration message and returns a quoted string.
CALibration:VOLTage[:DATA] <numeric value>
This command can only be used after calibration is unse cured and the out put
state is ON. It enters a voltage value that you obtained by reading an external
meter. You must first select the minimum calibration level (
CAL:VOLT:LEV
MIN
) for the value being entered . You must then select the middle and
maximum calibration levels (
CAL:VOLT:LEV MID
and
CAL:VOLT:LEV MAX
)
for the value being entered. Three successive values must be selected and
entered. The power supply then computes new voltage calibration constants.
These constants are then stored in
non-volatile
memory.

Table of Contents

Other manuals for Agilent Technologies E3633A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3633A and is the answer not in the manual?

Agilent Technologies E3633A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3633A
CategoryPower Supply
LanguageEnglish

Related product manuals