Chapter 2 129
Front-Panel Key Reference
Meas Setup
Avg Mode 
Exp Repeat
Allows you to select the type of termination control used for the 
averaging function. This determines the averaging action after the 
specified number of measurements (average count) is reached. 
Avg Mode has no effect on measurements that are not averaged.
Exp  Each successive data acquisition after the average 
count is reached is exponentially weighted and 
combined with the existing average. Exponential 
averaging weights new data more than old data, which 
facilitates tracking of slow-changing signals.
Repeat  After reaching the average count, all previous result 
data is cleared and the average count is set back to 1. 
Key Access: 
Meas Setup
Max Hold 
On Off
Allows you to turn maximum hold trace feature On or Off for the 
measurement. Maximum hold displays and holds the maximum 
responses of a signal.
Key Access: 
Meas Setup
Occ BW % Pwr 
Allows you to change the percentage of the total input signal power 
used when determining the occupied bandwidth. The acceptable range 
for this parameter is from 10% to 99.99%.
Key Access: 
Meas Setup
OBW Span 
Allows you to specify the range of integration used in calculating the 
total power from which the percent occupied bandwidth is then 
calculated. The analyzer span will be set to the same value as the OBW 
Span for the measurement. OBW Span should be set to approximately 2 
times the expected occupied bandwidth result. Refer to Table 2-3.
Key Access: 
Meas Setup
Table 2-3  Occupied Bandwidth Default Settings for Radio Standards
Standard Format Device OBW 
Span
VBW/RBW
None N/A N/A 3 MHz
10/1
1
IS95  N/A BTS/MS 2.0 MHz 100 kHz/
10 kHz
NADC  N/A BTS/MS 80 kHz 10 kHz/
1 kHz
W-CDMA  3GPP BTS/MS 6.0 MHz 1 MHz/
100 kHz
cdma2000  SR1 BTS/MS 2.0 MHz 100 kHz/
10 kHz