Chapter 2 145
Front-Panel Key Reference
Meas Setup
Sweep Time
Auto Man
Allows you to set the sweep time for the selected range.
When
Auto is selected, the analyzer determines the
optimum setting, while
Man allows you to enter the
setting.
Key Access:
Meas Setup, Range Table
Range
On Off
Allows you to select from the 20 ranges in the range
table. When a range has been selected, the values of the
range table parameters will appear on the
Range Table
menu keys. If
Range is set to On, the selected range
values will be used in your measurement. If
Range is
set to
Off, the selected range values will not be used.
The range parameters are listed in the table above.
Key Access:
Meas Setup, Range Table, More
Absolute Limit
Allows you to determine the limit above which spurs
will report a failure. Any spur located above the current
setting of
Peak Excursn will be added to the results
table. An F (in red on analyzers with a color display)
will be appended to the amplitude value of the spur if
the measured amplitude of the spur is above the limit
set with
Absolute Limit.
Key Access:
Meas Setup, Range Table, More
Peak Excursn
Sets the minimum amplitude variation of signals that
can be identified as peaks. For example, if a value of 6
dB is selected, peaks that rise and fall more than 6 dB
above the peak threshold value are identified.
Key Access:
Meas Setup, Range Table, More
Pk Threshold
Sets the minimum amplitude of signals that can be
identified as peaks. For example, if a value of –90 dBm
is selected, only peaks that rise and fall more than the
peak excursn value, which is above –90 dBm, are
identified.
Key Access:
Meas Setup, Range Table, More
Meas Type
Examine Full
Allows you to specify the measurement type, either Examine or Full.
This parameter is coupled to the average mode. If the measurement
type selected is
Examine, the measurement sets Avg Mode to Exp. When
Full is selected as the measurement type, the measurement sets Avg
Mode
to Rep. The behavior of each measurement type is described
below.

Table of Contents

Other manuals for Agilent Technologies E4402B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4402B and is the answer not in the manual?

Agilent Technologies E4402B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4402B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals