128 Appendix B
SCPI Command Reference
:CALC{1-4}:TRAN:TIME:CENT
:CALC{1-4}:TRAN:TIME:CENT
Syntax :CALCulate{[1]|2|3|4}[:SELected]:TRANsform:TIME:CENTer <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:TRANsform:TIME:CENTer?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4), sets the center value
used for the transformation function of the fault location function.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use 10 OUTPUT 717;":CALC1:TRAN:TIME:CENT 1E-8"
20 OUTPUT 717;":CALC1:TRAN:TIME:CENT?"
30 ENTER 717;A
Related commands :CALC{1-4}:TRAN:TIME:SPAN on page 134
:CALC{1-4}:TRAN:TIME:STAT on page 136
:CALC{1-4}:TRAN:DIST:CENT on page 113
Equivalent key [Analysis] - Fault Location - Center
<numeric>
Description Center value
Range Varies depending on the frequency span and the number of
points.
Preset value 0
Unit s (second)