Appendix B 145
SCPI Command Reference
:CALC{1-4}:SRL:CONN{1-2}:IMM
:CALC{1-4}:SRL:CONN{1-2}:IMM
Syntax :CALCulate{[1]|2|3|4}:SRL:CONNector{[1]|2}:IMMediate
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4), measures the
terminated cable connected to the specified port and automatically sets the connector
length and capacitance values for connector mismatch compensation. (No query)
Related commands :CALC{1-4}:TRAN:DIST on page 112
:CALC{1-4}:SRL:CONN{1-2}:IMP? on page 145
:CALC{1-4}:SRL:CONN{1-2}:LENG on page 146
Equivalent key [Analysis] - SRL - Portx Connector - Measure Connector
:CALC{1-4}:SRL:CONN{1-2}:IMP?
Syntax :CALCulate{[1]|2|3|4}:SRL:CONNector{[1]|2}:IMPedance?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4), reads out the average
cable impedance used for the SRL calculation. (Query only)
Query response {numeric}<newline><^END>
Example of use 10 OUTPUT 717;":CALC1:SRL:CONN1:IMP?"
20 ENTER 717;A
Related commands :CALC{1-4}:TRAN:DIST on page 112
:CALC{1-4}:SRL:CONN{1-2}:IMM on page 145
:CALC{1-4}:SRL:CONN{1-2}:LENG on page 146
Equivalent key None