6-28
Error Messages
ESG-D Series Signal Generator Instrument-Specific Error Messages
Settings conflict;Internal burst cannot be on with pulse
modulation.
Indicates that there is a hardware conflict between internal burst and pulse modulation.
The most recently requested modulation will be turned on, the previous modulation will
be turned off.
Settings conflict;Pattern repeat is changed to continuous because
data source is external.
Indicates that, while in non-bursted data generation, Pattern Repeat was in Single mode
and data source was selected to be External. For non-bursted data generation using an
external data source, Pattern Repeat must be in Continuous mode. To continue data
transmission, Pattern repeat has been changed to Continuous mode.
Settings conflict;Uplink & Downlink setting conflict. All
selected Uplink timeslots are now set to Downlink.
Indicates that an Uplink timeslot type has been selected while another timeslot(s) has
Downlink selected. In NADC and PDC digital modulation for bursted data, only either
Uplink (base to mobile) or Downlink (mobile to base) Traffic Channel type can be
selected for all timeslots at any one moment.To continue data transmission, all Uplink
timeslots have been changed to the Downlink configuration.
Settings conflict; Downlink & Uplink setting conflict. All
selected Downlink timeslots are now set to Uplink.
Indicates that an Downlink timeslot type has been selected while another timeslot(s) has
Uplink selected. In NADC and PDC digital modulation for bursted data, only either
Uplink (base to mobile) or Downlink (mobile to base) Traffic Channel type can be
selected for all timeslots at any one moment.To continue data transmission, all
Downlink timeslots have been changed to the Uplink configuration.
Settings conflict; Pulse period set less than pulse width. Pulse
width changed to match period value.
Indicates that a pulse period has been entered which is smaller than the pulse width. The
instrument automatically adjusts the pulse period to match the pulse width.
Settings conflict; PDC and NADC Custom protocols are now uplink
protocols, so the timeslots with Custom selected were changed to
Downlink Custom.
Indicates that, because PDC or NADC has downlink protocols selected and also has
Custom in one or more timeslots, the timeslots with Custom are set to Downlink
Custom. Otherwise, the timeslots with Custom are set to Uplink Custom. Downlink
Custom has its own instrument state information for the Data type and Fix 4 value.