6-40
Error Messages
ESG-D Series Signal Generator Instrument-Specific Error Messages
installed option, uninstall all options and then reinstall the correct options. If the error
persists, contact the factory.
Configuration error; Installed option boards do not match
configuration information.
This indicates that the option boards have not been properly installed in the instrument.
Verify that the correct option boards have been installed in the correct slots. Reinstall the
correct option. If the error persists, contact the factory.
Configuration error; Invalid Data Generator memory configuration.
This indicates that the memory configuration for an option board does not match the
known memory limits of any supported option board. If this error has occurred as the
result of a customer-installed option, uninstall all options and then reinstall the correct
options. If the error persists, contact the factory.
Configuration error; Invalid option board configuration.
This indicates that an invalid combination of option boards has been configured. If this
error has occurred as the result of a customer-installed option, uninstall all options and
then reinstall the correct options. If the error persists, contact the factory.
700
State Save Recall Error;Recall aborted. Unable to recall the
state from non-volatile memory.
This indicates that the state file was not readable, so the recall was aborted. If state file
exists, delete explicitly using the memory catalog.
State Save Recall Error;Recalled state has a bad checksum. No
instrument state change.
This indicates that the state file was corrupt or out-of-date, so the recall was ignored. If
state file exists, delete explicitly using the memory catalog.
State Save Recall Error;Recall data different from FW revision.
No instrument state change.
Indicates that an attempt was made to recall a state that was saved with an incompatible
version of the instrument firmware. This typically occurs when a state file is copied from
an instrument with a newer version of firmware to an instrument with an older version of
firmware. Newer versions of instrument firmware can read older state files.
State Save Recall Error;Recall non-saved state register. Recall
ignored.
Indicates that a recall was attempted for a state register that is unused. If state file exists,
delete explicitly using catalog.

Table of Contents

Other manuals for Agilent Technologies ESG-D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ESG-D Series and is the answer not in the manual?

Agilent Technologies ESG-D Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelESG-D Series
CategoryPortable Generator
LanguageEnglish

Related product manuals