1200 Series DAD and MWD User Manual 105
Maintenance 8
Note:
The label attached to the flow cell provides
information on part number, path length, volume and
maximum pressure.
If you want to replace flow cell parts, see
“Maintenance of Standard, Semi-Micro or Micro
Flow Cell” on page 106, “Maintenance of High
Pressure Flow Cell” on page 111 or “Nano Flow Cell
- Replacing or Cleaning” on page 123.
6 Insert the flow cell while pressing the flow cell holder.
7 Insert the flow cell capillaries into the union holder (top is
inlet, bottom is outlet). Tighten the thumb screw and
Reconnect the waste tubing (bottom) to the union.
Note:
To check for leaks, establish a flow and observe the
flow cell (outside of the cell compartment) and all
capillary connections.
Next steps:
8 Perform a “Wavelength Verification and Recalibration” on page 138 or a “Holmium Oxide Test” on page 139 to check the
correct positioning of the flow cell.
9 Replace the front cover.
inlet
outlet
thumbscrew
waste