1200 Series DAD and MWD User Manual 101
Maintenance 8
Exchanging a Lamp
When required If noise or drift exceeds application limits or lamp does not ignite
Tools required Screwdriver POZI 1 PT3
Parts required Longlife Deuterium lamp 2140-0813 (without black cover)
Longlife Deuterium lamp 5181-1530 (with black cover)
Tungsten lamp G1103-60001
Preparations Turn the lamp(s) off.
CAUTION
If the detector has been in use, the lamp may be hot. If so, wait five minutes until lamp
cools down.
1 Press the release buttons and remove the front cover to
have access to the lamp area.
2 Disconnect lamp from the connector and unscrew the
lamp and remove the lamp. Do not touch the glass bulb
with your fingers.