1200 Series DAD and MWD User Manual 71
How to optimize the Detector 5
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Flow Cell Path Length
Lambert-Beer’s law shows a linear relationship between the flow cell path
length and absorbance.
where
T is the transmission, defined as the quotient of the intensity of the
transmitted light I divided by the intensity of the incident light, I
0
,
ε is the extinction coefficient, which is a characteristic of a given substance
under a precisely-defined set of conditions of wavelength, solvent,
temperature and other parameters,
C is the concentration of the absorbing species (usually in g/l or mg/l), and
d is the path length of the cell used for the measurement.
Therefore, flow cells with longer path lengths yield higher signals. Although
noise usually increases little with increasing path length, there is a gain in
signal-to-noise ratio. For example, in
Figure 26 the noise increased by less
than 10 % but a 70 % increase in signal intensity was achieved by increasing
the path length from 6 mm to 10 mm.
When increasing the path length, the cell volume usually increases in our
example from 5 13 µl. Typically, this causes more peak dispersion. As
Figure 26 demonstrates, this did not affect the resolution in the gradient
separation in our example.
As a rule-of-thumb the flow cell volume should be about 1/3 of the peak
volume at half height. To determine the volume of your peaks, take the peak
width as reported in the integration results multiply it by the flow rate and
divide it by 3).
Absorbance Tlog
I
0
I
----log ε Cd⋅⋅===

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1365B and is the answer not in the manual?

Agilent Technologies G1365B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1365B
CategorySecurity Sensors
LanguageEnglish

Related product manuals