134 1200 Series DAD and MWD User Manual
8Maintenance
Replacing the Detector’s Firmware
The installation of older firmware might be necessary:
• to keep all systems on the same (validated) revision, or
• if third part control software requires a special version.
To upgrade/downgrade the detector’s firmware the following steps have to be
performed:
1 Download the module’s firmware, the LAN/RS-232 FW Update Tool Version
2.10 or above and the documentation from the Agilent web
http://www.chem.agilent.com/scripts/cag_firmware.asp.
2 Load the firmware into the detector as described in the documentation.
When required If new version solves problems of currently installed version or after exchange
of the detector main board (VWM) the version on board is older than previous
installed one.
Tools required LAN/RS-232 Firmware Update Tool, or
Instant Pilot G4208A or Control Module G1323B
Parts required Firmware, tools and documentation from Agilent web site
Preparations Read update documentation provided with the Firmware Update Tool.
NOTE
The G1315B DAD and the G1365B MWD require firmware revision A.04.x or above (main
and resident).