106 1200 Series DAD and MWD User Manual
8Maintenance
Maintenance of Standard, Semi-Micro or Micro Flow Cell
When required If the flow cell needs repair due to leaks or contaminations (reduced light
throughput)
Tools required Two 1/4 inch wrenches for capillary connections
hexagonal key 4 mm
Tooth picks
Parts required For parts, see
“Standard Flow Cell on page 146, Semi-Micro Flow Cell on
page 148 or
“Micro Flow Cell” on page 150.
Preparations Turn the flow off.
Remove the front cover.
Remove the flow cell, see Exchanging a Flow Cell” on page 103.
NOTE
The gaskets used in the standard and semi-micro/micro flow cell are different.