140 1200 Series DAD and MWD User Manual
8Maintenance
Suggested Actions
Ensure the flow cell is filled with water.
Recalibrate (see Wavelength Verification and Recalibration” on page 138)
and repeat the test.
Run the cell test (see “Cell Test (ChemStation only)” on page 141). If the test
fails, exchange the flow cell windows.
Clean optical components with alcohol and lint-free cloth.
Exchange the UV lamp.
Figure 44 Holmium Oxide Test Results (report)