System Maintenance
NovoCyte Quanteon Flow Cytometer Operator's Guide 107
6 System Maintenance
This chapter describes the procedures to maintain NovoCyte Quanteon flow
cytometer, including scheduled and unscheduled maintenance.
It is important to conduct the instrument maintenance on a regular basis in order
to achieve the optimal performance for the instrument and obtain reliable data.
Agilent flow cytometer incorporates automatic procedures for maintenances as
much as possible in order to bring conveniences to the users and save them
valuable time. Most maintenance procedures will be automatically executed,
eliminating the time-consuming manual maintenances.
Scheduled Maintenance
The following maintenance should be conducted on regular basis.
Table 9 Scheduled Maintenance and Recommended Frequency
Maintenance
Recommended Maintenance Frequency
Daily Start-up
Daily. Automatic system cleaning during start-up.
Daily Shutdown
Daily. Automatic system cleaning and
decontamination during shutdown.
Add instrument reagents (NovoFlow sheath fluid,
NovoRinse solution, and NovoClean solution)
As needed. Check the instrument reagent volume
in the containers before each experiment and add
more reagents when necessary.
Empty Waste
As needed. Check waste volume before each
experiment and empty waste container when
necessary.
Preventive Maintenance
Clean the sample injection probe (SIP)
Clean the SIP cleaning apparatus
Unclog the flow cell
Sterilize instrument reagent containers
Every month.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies NovoCyte Quanteon and is the answer not in the manual?

Agilent Technologies NovoCyte Quanteon Specifications

General IconGeneral
BrandAgilent Technologies
ModelNovoCyte Quanteon
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals