System Maintenance
130 NovoCyte Quanteon Flow Cytometer Operator's Guide
Calibrate NovoCyte Fluidics Station II
NovoCyte Fluidics Station II is calibrated at the factory and should be ready to use
when connected to the instrument after installation. However, in some cases when
there is a need to recalibrate the fluidics station after shipment and movement
(e.g. when a false warning about the liquid level in the reagent containers is
observed), follow the procedures below to recalibrate the fluidics station.
NOTE
Calibrate Fluidics Station function is only available when Calibrate Fluidics Station
privilege is enabled. Refer to NovoExpress Software Guide for instructions to enable
this privilege.
1 Click Instrument > Operation > Calibrate Fluidics Station in the NovoExpress
software.
Figure 110. Calibrate Fluidics Station Function in NovoExpress
2 Follow the instruction in the software, remove the NovoFlow, NovoRinse,
NovoClean, and waste containers from the fluidics station. Click the OK
button.
Figure 111. Software Message Instructing to Start Fluidics Station Calibration
3 The software will prompt a dialog box indicating the fluidics station calibration
is successful. Follow the instructions to place the NovoFlow, NovoRinse,
NovoClean, and waste containers back into the fluidics station. Click the OK
button.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies NovoCyte Quanteon and is the answer not in the manual?

Agilent Technologies NovoCyte Quanteon Specifications

General IconGeneral
BrandAgilent Technologies
ModelNovoCyte Quanteon
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals