EasyManuals Logo

Agilent Technologies NovoCyte Quanteon User Manual

Agilent Technologies NovoCyte Quanteon
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
NovoCyte Quanteon Technical Specifications
NovoCyte Quanteon Flow Cytometer Operator's Guide 143
Appendix B: NovoCyte Quanteon Technical
Specifications
This chapter provides technical specifications of NovoCyte Quanteon flow
cytometer and associated standard accessories.
NovoCyte Quanteon Technical Specifications
Table 15 NovoCyte Quanteon Technical Specifications
1
Optics
Laser
Solid state laser with on-board thermal-electric
cooling and guaranteed thermal stability and
beam quality
Laser Power
100 mW
Laser Beam Profile
10 x 60 µm elliptical beam
Optical Alignment Procedure
Fixed, no operator alignment required
Fluorescence Detection
Silicon Photomultiplier (SiPM) with high photon
detection efficiency; Individual photodetector for
each channel
FSC/SSC Sensitivity
FSC: 0.4 µm; SSC: 0.1 µm
Fluorescence Resolution
<3% CV for CEN
Optical Filters
User exchangeable, “Smart” filter automatically
read by the system
Cell Size
0.1 - 50 µm
Fluidics
Flow Cell
170 x 290 µm rectangular quartz flow cell
Sample Acquisition Rate
100,000 events/second
Sample Delivery
Positive-displacement syringe pump enabling
direct absolute count without the need for
reference counting beads.
Volumetric Absolute Count
Precision
< 5%

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies NovoCyte Quanteon and is the answer not in the manual?

Agilent Technologies NovoCyte Quanteon Specifications

General IconGeneral
BrandAgilent Technologies
ModelNovoCyte Quanteon
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals