System Maintenance
110 NovoCyte Quanteon Flow Cytometer Operator's Guide
WARNING
Sample injection probe directly contacts the biological samples which may
be biohazardous. Wear gloves when handling the sample injection probe.
CAUTION
Sample injection probe is fragile and may be damaged if handled
improperly. When cleaning the sample injection probe, please be careful
not to deform or bend it.
6 Click OK button after cleaning.
Figure 84. Software Message Instructing the Cleaning Procedures
7 Place a tube of Milli-Q or 0.2 µm filtered DI water or NovoRinse solution (1X) in
the tube holder. The instrument will shut down automatically after clicking OK
button.