NovoCyte Quanteon Technical Specifications
NovoCyte Quanteon Flow Cytometer Operator's Guide 145
Table 15 NovoCyte Quanteon Technical Specifications
1
Administrative creation of individual user
accounts and user groups with privilege control.
Configurable roles for individual users. User
operation time tracking.
Automatic daily quality control with Agilent Flow
Cytometer QC Particles. Levi-Jennings Plot.
Each instrument is verified for assembly
integrity and performance to specifications.
12 X 75 mm tube, 1.5 mL or 2.0 mL tube with
Agilent tube holder.
Automatic Sample Loading
with NovoSampler Q
Agilent Flow Cytometer 40 tube rack for 12 x 75
mm tube, 24-well, 48-well, 96-well (flat, V-, U-
bottom), 384-well microtiter plates, Agilent Flow
Cytometer 24 tube cooling rack for 12 x 75 mm
tube, Agilent Flow Cytometer 96-well (flat, U-, V-
bottom) plate cooling box.
Instrument Dimension
(W × D × H)
24.4 × 18.1 × 18.8 in (62.0 × 46 ×48.0 cm)
33.5 × 18.1 × 18.8 in (85.0 × 46.0 ×48.0 cm) with
NovoSampler Q
104.0 lb (47.0 kg)
133.0 lb (60.0 kg) with NovoSampler Q
Temperature:15-32℃, Relative Humidity: 80%
maximum, Atmosphere Range: 76-106 kPa
Temperature:1-40℃, Relative Humidity: 80%
maximum
Class I Laboratory Equipment
Pollution degree of the
intended environment
Degree of Ingress protection
1
Specification and performance claims would be based on certain specific experiment conditions.