NovoCyte Quanteon Technical Specifications
NovoCyte Quanteon Flow Cytometer Operator's Guide 145
Table 15 NovoCyte Quanteon Technical Specifications
1
User Management
Administrative creation of individual user
accounts and user groups with privilege control.
Configurable roles for individual users. User
operation time tracking.
Quality Control
Instrument Daily QC
Automatic daily quality control with Agilent Flow
Cytometer QC Particles. Levi-Jennings Plot.
Production QC
Each instrument is verified for assembly
integrity and performance to specifications.
Sampling
Manual Sample Loading
12 X 75 mm tube, 1.5 mL or 2.0 mL tube with
Agilent tube holder.
Automatic Sample Loading
with NovoSampler Q
Agilent Flow Cytometer 40 tube rack for 12 x 75
mm tube, 24-well, 48-well, 96-well (flat, V-, U-
bottom), 384-well microtiter plates, Agilent Flow
Cytometer 24 tube cooling rack for 12 x 75 mm
tube, Agilent Flow Cytometer 96-well (flat, U-, V-
bottom) plate cooling box.
Operating
Conditions
Instrument Dimension
(W × D × H)
24.4 × 18.1 × 18.8 in (62.0 × 46 ×48.0 cm)
33.5 × 18.1 × 18.8 in (85.0 × 46.0 ×48.0 cm) with
NovoSampler Q
Instrument Weight
104.0 lb (47.0 kg)
133.0 lb (60.0 kg) with NovoSampler Q
Operating Environment
Temperature:15-32, Relative Humidity: 80%
maximum, Atmosphere Range: 76-106 kPa
Storage Environment
Temperature:1-40, Relative Humidity: 80%
maximum
Power Requirements
100-240 VAC, 50-60 Hz
Regulatory
Compliance
Indoor and Outdoor Use
Indoor use only
Overvoltage Category
Class II
Equipment Class
Class I Laboratory Equipment
Pollution degree of the
intended environment
2
Degree of Ingress protection
IPXO
1