If the result of certain detection channels shows Acceptable or Failed as below,
make sure the QC particles are properly prepared. Run the QC particles once again
after the correct action has been taken. Ensure there is at least 400 μL sample
remaining in the sample tube. Click Repeat Test to repeat the QC test. The
photodetector gain will be automatically re-adjusted.