30
Navigating and Customizing Your Flow Controller
Choosing Device Engineering Units from the Basic Conguration Menu
Changing device engineering units alters both the display and the data frame.
First choose the parameter whose unit you want to change, and then select
your desired engineering unit, conrming the change on the last screen. If
your controller has been congured with a ow totalizer, this screen will also
include units for totalized volumetric and mass ow, plus elapsed time.
Basic Conguration Menu
The Basic Conguration Menu contains options for choosing the gas
calibration, device engineering units and STP/NTP mass ow references.
SETCANCEL
PRESSING SET WILL
AFFECT DISPLAY
AND SERIAL VALUES.
VERIFY CONNECTED
SERIAL DEVICES
EXPECT THE CHANGE.
Menu | Basic Conguration
MAIN
BACK
STP/
NTP
DEVICE
UNITS
GAS
Air
MAIN Exits to the Main Display
(page 19).
BACK Returns to the top-level
Menu (page 24).
DEVICE UNITS Changes device engineering
units for any parameter:
Mass Flow - Volumetric Flow -
Pressure - Temperature -
(Volu Totalizer - Mass Totalizer - Totalizer Time)
STP/NTP Denes standard
(STP) and normal (NTP)
temperature and pressure
conditions (page 35).
GAS Enters Gas Select and
COMPOSER menus (page 31).
SELECTBACK
DOWNUP
Totalizer Time
Mass Totalizer
Volu Totalizer
Pressure
Temperature
Volumetric Flow
Mass Flow
>