24
Navigating and Customizing Your Flow Controller
Menu
You can enter the menu system by pressing the MENU button from the Main Display.
Taring Your Flow Controller
Taring is an important practice that ensures that your ow controller is
providing the most accurate measurements possible. This function gives the
ow controller a good zero reference for ow measurements. For controllers
with a barometer, taring can also be used to align the internal absolute
pressure sensor with the barometric pressure reading.
How to Tare
When auto tare is -ON- your ow controller automatically
tares its ow rate whenever it has a zero setpoint for more
than 1.2 seconds. For manual tares, follow these steps:
1. Ensure that nothing is owing through the device,
usually by giving the controller a zero setpoint.
2. MENU > TARE> TARE FLOW. Flow tares should occur at the
expected process pressure, as long as there is no ow.
3. MENU > TARE> TARE PRESS Absolute pressure tares must be
done with the controller open to atmosphere.
When to tare
• After signicant changes in temperature or pressure.
• After installing the controller in a dierent orientation.
• After dropping or bumping the ow controller.
Menu
MAIN
ADV
SETUP
BASIC
CONFIG
TARESABOUTCONTROL
MAIN Exits to the Main Display
(page 19).
ADV SETUP Enters the Advanced
Setup Menu (page 36).
BASIC CONFIG Enters the Basic
Conguration Menu (page 30).
TARES Enter the Tares Menu
ABOUT Enters the About Menu
(page 29).
AUTO
TARE
-ON-
BACK MAIN
TARE
FLOW
TARE
PRESS
TARECANCEL
PRESS TARE WHEN
VENTED TO AMBIENT
WITH NO FLOW.
Current pressure
offset:
+0.8 mmHgA
TARE PRESS
TARE FLOW
TARECANCEL
ENSURE NO FLOW
BEFORE
PRESSING TARE
CONTROL Enters the Control
Menu (page 25).