2-16 | Model 931S / Model 932S UV Analyzers
ATEX- and IECEx- certified Model 931S and 932S Analyzers are marked
with the label shown below.
Power Dissipation
<310W
<310W
Input
240VAC 50/60Hz 1.5A max.
120VAC 50/60Hz 3A max.
Certificate Nos.: KEMA 02ATEX2247 X, IECEx DEK12.0035X
Western Research Model 93X Analyzer
Year: _______ Serial No.: __________________
Cable Entries
M25 X 1.5
II 2 G
0344
Caution
Do not open when an explosive gas atmosphere is present.
Use Stainless Steel Fasteners with yield strength ≥450 MPa (65,300 psi).
Ex db IIB+H2 T3 Gb (T amb. 0°C...50°C)