22 User Guide EN Dansensor® CheckMate 3
P/N 300042-I
05/2019
Measuring modes
Four different measuring modes are available:
Manual spot Press   key to start the measuring.
The measuring stops automatically after the set-up product measuring 
time.
Auto spot The measuring starts automatically when the measured gas is different 
from 20.9% O
2
 +/- 1.5%.
When the gas measuring is stable, it stops automatically.
To stop measuring manually press   key.
Intermittent Press   key to start the measuring.
The device will now perform continuous measurements at fixed intervals 
until it is stopped again by pressing the   key.
The measurement time for each measurement and the time interval 
between measurements are set up in the product set-up menu.
Continuous Press   key to start the measuring.
The unit will now perform continuous measurements until it is stopped 
again by pressing the   key.
NOTE! Even though the pump is high quality, it is recommended to stop 
Continuous/Auto spot measurements when not used for longer periods.
The various measuring mode parameters are set up in the product set-up menu.
See "Edit product" on page 33 for details.
NOTE! During intermittent and continuous measurement, the ambient pressure 
can not be measured.
Performing intermittent and continuous measurements during fluctuating 
ambient pressure will impact the accuracy.
Stop measurement and place sample needle at ambient pressure for a couple of 
minutes to obtain correct ambient pressure.