Calibration | 5-21
Set Cal Timers - Set Auto Timers
Set the frequencies for automatic calibrations or verifications, including
the time when the first calibration or verification should begin. You have
the option of setting the cycle in either hours or days.
If you schedule a verification and calibration at the same time, the
verification will be performed first.
1. From the Set Cal Timers menu, select Set Auto Timers.
2. Select Auto Cal Cycle if defining an automatic calibration cycle. Select
Auto Verify Cycle if defining an automatic verification cycle.
3. Select Set Cycle in DAYS or Set Cycle in HOURS, depending on
whether you want to set the frequency of calibrations or verifications
in days or hours.
If you are setting the auto timers for the first time, the Series 2000
software will automatically prompt you to enter the current time and
date.
You will be prompted to enter the frequency for the calibration or verifica-
tion cycle (either in days or hours, depending on what you selected:
Cal Cycle = 01 days
New Value = xx
4. Enter the number of days (or hours) for the automatic calibration or
verify cycle. As you enter the values, they overwrite the XXs on the
display. To enter 1 day or 1 hour, enter 01; otherwise, the system will
assume you are entering 10 days (or hours).
You will be prompted to enter the time of day when the first calibra-
tion of the cycle should begin:
Cal time is YY:YY
Enter new time Hr:Mn
5. Enter the time you want the first calibration or verification to begin,
then press Enter. The system uses military time, so to enter 4:00 p.m.,
you would type 16:00. If, for example, you set the cycle frequency at
8 hours, and specify the first calibration to take place at 4:00 p.m., the
next automatic calibration or verification will take place at midnight.
NOTE
NOTE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Thermox Series 2000 WDG-IVC and is the answer not in the manual?

Ametek Thermox Series 2000 WDG-IVC Specifications

General IconGeneral
BrandAmetek
ModelThermox Series 2000 WDG-IVC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals