6-12   |   Thermox TM2000 Oxygen Analyzer
Remote Calibration Unit (RCU) problem
Problems with the RCU are usually as follows:
  Vacuum check
  Plumbing leaks
  Solenoids not energizing
Vacuum Check
Perform a vacuum check as follows:
1.  Connect a digital volt meter across the connections labeled 1 and 2 on 
the sensor board. Observe polarity (pin 1 - positive, pin 2 - negative).
2.  Connect the lowest calibration gas value that you will encounter dur-
ing daily operations to the sample gas inlet of the analyzer. Allow gas 
to stabilize then measure the cell millivolt signal.
3.  Connect the vacuum port side of the pressure gauge (for example, a 
Magnehelic gauge) to the T-fitting.
4.  Connect the sample gas outlet to the T-fitting.
5.  Connect an aspirator to the T-fitting.
6.  Connect the air supply to the aspirator.
7.  Adjust sample gas and air supply flow valves until the flow rate is 150 
ml/min and the pressure gauge is slightly greater that 25” H
2
O pres-
sure.
8.  Measure the cell millivolt signal with the digital volt meter. If the cell 
millivolt signal stays the same or increases, the analyzer has passed 
the vacuum test. If the cell millivolt signal decreases, troubleshoot 
to find the leak, then repeat Steps 1 through 7. Possible source of 
leaks include: 1) the sample gas line, 2) compression fittings, and 3) a 
cracked cell.