6-16 | Thermox TM2000 Oxygen Analyzer
If your analyzer doesn’t respond properly to the calibration gas, then it
may indicate a problem with the cell. Before replacing the cell, put the
analyzer through a primary calibration. If this does not work, check the
following:
Check the proper flow rate and pressure of the gases.
Check for leaks or plugged plumbing in the sensor.
Check that the cell is hot enough.
If this does not solve the problem, replace the cell and run another prima-
ry calibration and a regular calibration after the primary calibration.