Serial Communications | A-13
Figure A-7.
I_out ags variable table.
Bit Description Read/Write * Format
0 track/hold output #1 during cal 1=track, 0 = hold
1 track/hold output #1 1 = track, 0 = hold
2 current mode select for output #1 1 = 0 - 20 mA 0 = 4 - 20 mA
3 Reserved N/A
4 track/hold output #2 during verify 1 = track, 0 = hold
5 track/hold output #2 during verify 1 = track, 0 = hold
6 current mode select for output #2 All Read/Write 1 = 0 - 20 mA 0 = 4 - 20 mA
7 Reserved N/A
8 track/hold output #3 during cal 1 = track, 0 = hold
9 track/hold output #3 during verify 1 = track, 0 = hold
10 current mode select for output #3 1 = 0 - 20 mA 0 = 4 - 20 mA
11 Reserved N/A
12 track/hold output #4 during cal 1 = track, 0 = hold
13 track/hold output #4 during verify 1 = track, 0 = hold
14 current mode select for output #4 1 = 0 - 20 mA 0 = 4 - 20 mA
15 Reserved N/A