EasyManuals Logo

Ametek Thermox TM2000 User Manual

Ametek Thermox TM2000
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
4-14 | Thermox TM2000 Oxygen Analyzer
primary calibration, then the furnace temperature set point cannot be in-
creased any further to prevent damage to the cell housing and/or furnace
assembly.
Failure limits for standard calibration -
+/- 10 mVs of the anticipated mVs for the gas concentration applied (air 0
mV expected).
+/- 5 mVs of the anticipated mVs for the gas concentration applied (2%
O2 applied, 44 mV expected at 615°C cell temp) as referenced to the actual
span gas mV reading.
T/C ratio adjustment limits for Primary Calibration -
For a cell operating temperature of 615°C, the T/C ratio low limit is 0.7
and the high limit is 1.3, which corresponds to a thermocouple indicated
temperature range of 473°C to 878°C.
If the analyzer is configured for a cell operating temperature of 615°C, the
highest possible operating temperature for the analyzer at after a primary
calibration is 878°C.
This menu option allows you to perform a primary calibration on the
zirconium oxide oxygen sensor. Perform a primary calibration on your
system rather than a regular calibration (from the Calibrate Key) in the
following cases:
When you replace your zirconium oxide oxygen cell.
When your system fails a regular calibration.
CELL TEMP °C MAX T/C RATIO MIN T/C RATIO
615 1.3 0.7
650 1.24 0.74
695 1.22 0.777
824 1.139 0.861
790 same as 824°c

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Thermox TM2000 and is the answer not in the manual?

Ametek Thermox TM2000 Specifications

General IconGeneral
BrandAmetek
ModelThermox TM2000
CategoryAnalytical Instruments
LanguageEnglish