A-3 Test Records Appendix A — Test Records
A-10 PN: 13000-00164 Rev. K ML248xx, ML249xA MM
ML248xx, ML249xA Firmware Revision: Operator:
Serial Number: Date: Options:
Calibrator Power Level Uncertainty
Equations for Uncertainty Calculations
Standard Uncertainty Calculation
Source Uncertainty Calculations
Standard Uncertainty X
unc
1
Divisor()
-------------------------
Uncertainty()SensitivityC
ix
()××=
V
comp
Unc
1
3
-------
0.00003 V
comp
× 0.000004 10V×+()
2 V
1
V
0
()
4 R× EE× M×
---------------------------------------


××=
V
0
Unc
1
3
-------
2 0.00003 V
comp
× 0.0000004 10V×+()
2
×()
2 V
comp
V
0
()
4 R× EE× M×
---------------------------------------


××=
V
1
Unc
1
3
-------
2 0.00003 V
comp
0.000004 10V×+×()
2
()
2 V
comp
V
1
()
4 R× EE× M×()
--------------------------------------------


××=
R
Unc
1
3
-------
0.00006 R× 0.000002 1000()×+()
2V
comp
V
1
V
0
()V
0
2
V
1
2
+
4 R
2
× EE× M×
-----------------------------------------------------------------------------



××=
EE
Unc
1
1
---
U
e
2
-------


2V
comp
V
1
V
0
()V
0
2
V
1
2
+
4 R× EE
2
× M×
-----------------------------------------------------------------------------



××=
M
Unc
1
2
-------
2 Γ
s
×Γ
d
×()
2V
comp
V
1
V
0
()V
0
2
V
1
2
+
4 R× EE× M
2
×
-----------------------------------------------------------------------------



××=
CR
Unc
1
1
---
0.1%()0.001W()×× 0.000001W==
Combined Uncertainty V
comp
Unc
2
V
1
Unc
2
V
0
Unc
2
R
Unc
2
EE
Unc
2
M
Unc
2
CR
Unc
2
++++++=
K 2=()2 Combined Uncertainty 1000 (95% level of confidence)××=