Chapter 3 Prepare the Reactions
Prepare the Reaction Plate
51
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Standard Curve
Experiments
Notes
Prepare the Reaction Plate
Prepare the reactions for each replicate group, then transfer them to the reaction plate
using the plate layout displayed in the 7500 software.
About the
Example
Experiment
For the standard curve example experiment:
•A MicroAmp
®
Optical 96-Well Reaction Plate is used.
The reaction volume is 25 µL/well.
The reaction plate contains:
6 Unknown wells
15 Standard wells
3 Negative control wells
72 Empty wells
The plate layout that the 7500 software automatically generates is used: