Chapter 4 Generating Data from RQ Plates – 7300 or Standard 7500 System
Preparing the Reaction Plate
24 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Relative Quantification Getting Started Guide
Notes
STANDARD
STANDARD
STANDARD
Sample Experiment
Primers and probes for the example RQ experiment are obtained from the TaqMan Gene Expression Assays product line
and are provided as a 20 Gene Expression Assay Mix. The PCR master mix is prepared as follows:
Samples and endogenous controls are arranged on three plates as shown below. 50 µL of PCR master mix containing
cDNA are added to each well.
The reactions are kept on ice until the plates are loaded on the 7300/7500/7500 Fast system.
Reaction Component
µL/
Sample
µL/ 5 Reactions
§
§ 24 master mixes are prepared, one for each of 23 genes plus the endogenous control. Volume for five reactions (4 replicates plus extra)
to account for pipetting losses.
Final Concentration
TaqMa n
®
Universal PCR Master Mix (2) 25.0 125.0 1
20 Gene Expression Assay Mix
Contains forward and reverse primers and labeled probe.
2.5 12.5 1
cDNA sample 5.0 25.0 50 ng (for the 50-µL reaction)
Nuclease-free water 17.5 87.5
Total 50.0 250
GR2323
Regulus
Liver_96Plate.eps
GR2323
Endogenous
controls (GAPDH)
Liver
samples
GR2324
Regulus
Kidney_96Plate.eps
GR2324
Endogenous
controls (GAPDH)
Kidney
samples
GR2325
Regulus
Bladder_96Plate.eps
GR2325
Endogenous
controls (GAPDH)
Bladder
samples

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)