EasyManua.ls Logo

Applied Biosystems 7500 - Page 41

Applied Biosystems 7500
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 Generating Data from RQ Plates – 7300 or Standard 7500 System
Specifying Thermal Cycling Conditions and Starting the Run
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Relative Quantification Getting Started Guide 31
Notes
STANDARD
STANDARD
STANDARD
4
Thermal Cycling
Conditions for
One-Step RT-PCR
If you select the one-step RT-PCR method, cDNA generation and amplification take
place simultaneously at this point in the workflow.
The following table shows the thermal cycling conditions for one-step RT-PCR
experiments.
Note: Refer to the Online Help for instructions on modifying thermal cycling
parameters.
Times and Temperatures (One-step RT-PCR)
Initial Steps PCR (Each of 40 Cycles)
Reverse Transcription AmpliTaq
®
Gold DNA
Polymerase Activation
Melt Anneal/Extend
HOLD HOLD CYCLE
30 min @ 48 ° C 10 min @ 95 ° C 15 sec @ 95 ° C 1 min @ 60 ° C

Table of Contents

Other manuals for Applied Biosystems 7500

Related product manuals