Chapter 6 Analyzing Data in an RQ Study
Adjusting the Baseline and Threshold
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Relative Quantification Getting Started Guide 69
Notes
6
3. Set the baseline for the detector.
a. Under Analysis Settings, select Manual Baseline.
b. Enter values in the Start Cycle and End Cycle fields, ensuring that the
amplification curve growth begins at a cycle after the End Cycle value.
Note: After you change a baseline or threshold setting for a detector, the Analyze
button ( ) is enabled, indicating that you must reanalyze the data.
4. Set the threshold for the detector.
a. Under Analysis Settings, select Manual Ct.
b. Drag the threshold setting bar so the threshold is:
Above the background
Below the plateaued and linear regions of the amplification curve
Within the exponential phase of the amplification curve
The SDS Software 1.3.1 adjusts the threshold value and displays it in the
Threshold field after reanalyzing.
5. Repeat steps 2 through 4 to set the baseline and threshold values for all remaining
detectors in the study.
6. Click or select Analysis > Analyze to reanalyze the data using the adjusted
baseline and threshold values.
Click and drag the Threshold setting
to adjust the threshold. The bar turns
red, indicating that the threshold has
been changed.

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals