ACTA 4000 User Guide Printouts from ACTA
9836 4171 01 2008-12 109 (136)
8.9 Transducer calibration report
TRANSDUCER CALIBRATION REPORT
Transducer serial number Date Time
20529 March 28 2008 10:55.12
ACTA SERIAL NUMBER 1234567890
OTHER EQUIPMENT .................................................
.................................................
PREVIOUS CALIBRATION DATE Sept 08 2007
TRANSDUCER TYPE IRTT WITH ANGLE ENCODER
SENSITIVITY 2.0 mV/V
CALIBRATION INPUT No Load Cal point Cal value
(Nm) (mV/V) (Nm) (mV/V) (Nm)
0 -0.05 175 1.98 172.4
0 -0.02 175 1.95 177.7
0 -0.05 175 1.91 178.6
STANDARD DEVIATION 2.74
LINEARITY CHECK True value Measured value Unlinear
(Nm) (Nm) (mV/V) (%)
87 89.8 3,22 +1.59
132 130.3 -1,29 -0.97
STANDARD DEVIATION 1.32
OLD CALIBRATION VALUE 173.8 Nm
NEW CALIBRATION VALUE 176.1 Nm SAVED IN TRANSDUCER .... ....
YES NO
NEXT CALIBRATION DATE .................................................
ADDITIONAL INFORMATION .................................................
.................................................
.................................................
TEST PERFORMED BY
SIGNATURE
............................