Calibrating tools and equipment ACTA 4000 User Guide
94 (136) 2008-12 9836 4171 01
Non-controlled tool calibration
Use None to calibrate manual tools and tools without a display. See figure below.
Text on
display
Comment
X bar
Mean torque value.
Target
Target torque for tool.
Difference %
Difference in percent.
100⋅
−
=
Xbar
TargetXbar
%Difference
Soft key Comments
Store
Mark tightening operations as used in calibration. Save back-up tool question asked.
Printout question is asked.
Cancel
Close window.
Target
Update target torque for tool.
7.2 Calibrating Transducers with ACTA
ACTA has a built-in feature that allows calibration of transducers using the deadweight method.
When performing this task, the use of certified dead-weight equipment (arm and
weight) is needed. It is recommended that this task be carried out by Atlas Copco
service personnel or an approved laboratory.
In addition, if an Atlas Copco memory transducer is used and calibrated by means of the above method, it
is possible to download the new calibration value to the memory of the transducer directly from ACTA.
1. Place the transducer you wish to calibrate in the deadweight equipment.
2. Connect the transducer to ACTA.