E1070 Service Manual 125
3.5 Supervisor Menu
8. Toggle between Cntr and Print by using the TARE or UNITS key and press
ENTER when your choice is displayed…
If you choose Cntr, the overload or underload count will be displayed.
If you choose Print, you can choose the port to print through and press
ENTER. The counter value for overload or underload will be printed.
Example:
Overloads:
0 overloads
00:00 on 00-00-00
Underloads:
0 underloads
00:00 on 00-00-00
9. When you are done with the error log section, press the SELECT key
repeatedly until…
Err.Log is displayed.
10. Press the UNITS key…
S-Com is displayed. This is the SensorComm error log. See a sample
and the note below.
SensorComm Log:
Error #1 @ 15:00 on 12-17-03
11. Press the PRINT key…
Print is displayed.
12. Press the PRINT key to print the report.
Port 1 or Port 2 is displayed. Use this item to select which port to use for
printing the report.
13. Toggle between Port 1 and Port 2 using the TARE or UNITS key. Press the
ENTER key when the choice you want is displayed…
Display shows buSY briefly then returns to Print.
14. Press the UNITS key…
CLEAr is displayed. Use this to clear the log from memory.
15. Press PRINT to clear the log…
SurE ? is displayed.
See the section SensorComm Errors to see the list of error codes.
The S-COM log only appears if the scale is configured as a SensorComm scale. The
Ghost log will only appear if ghosting is enabled.